Ultracortex Mark IV

Development Period: January 2016 Through the Present

image

Complete Ultracortex Mark 4 kits are available for sale in the OpenBCI Online Store!

The Ultracortex is an open-source, 3D-printable headset intended to work with the OpenBCI system. It is capable of recording research-grade brain activity (EEG), muscle activity (EMG), and heart activity (ECG). The Ultracortex is under constant iteration. If you print/assemble your own Ultracortex, we would love to hear your feedback. Send us an email at contact@openbci.com or Tweet at us (@Ultracortex & @OpenBCI)!

Designers & Engineers:

THE COMPLETE ULTRACORTEX

Note: the part quantities listed below assume you are making an electrode holder for all 35 nodes of the Ultracortex Mark 4. In reality, you will likely have only 8 or 16 electrodes, depending on whether you are working with the OpenBCI Cyton Board (8 channels) or the OpenBCI 16-channel R&D Kit (16 channels). In general, more electrodes will distribute the downward scalp pressure, increasing comfort.

###3D-printed Parts:

The following parts are included in the Unassembled version of the Mark IV:

Non-printed Parts:

Note that unlike the Mark III, the electrodes and the electrode holders of the Mark IV are not designed to be 3D printed. The Spikey, Flat, and Comfort Units are all custom made and injection molded and can be purchased at the OpenBCI shop. If you need thse files for prototyping you can find them here.

The following parts are included in the Unassembled and the Print it Yourself versions of the Mark IV:

The following parts are not included with the purchase of any configuration of the Mark IV and should be purchased separately:

THE COMPLETE ULTRACORTEX (w/ PICTURES)

3D-printed Parts:

(1x) FRAME

  • FRAME (head circumference: small = 42-50cm, medium = 48-58cm, large = 58-65cm)
    • FRAME_FRONT (x1) — .STLs (small / medium / large )
    • FRAME_BACK (x1) — .STLs (small / medium / large)
    • You can also print the frame in quarters or print the full frame. The .stl files are found here.

image

(21x) MECH_PARTS

####(1x) BOARD_HOLDER

image

(1x) BOARD_COVER

image

image

Non-printed Parts:

Note that unlike the Mark III, the electrodes and the electrode holders of the Mark IV are not designed to be 3D printed. The Spikey, Flat, and Comfort Units are all custom made and injection molded and can be purchased at the OpenBCI shop. If you need thse files for prototyping you can find them here.

Suggested screws for fastening BOARD_MOUNT & OpenBCI Board

image

image

image

image

image

image

(1x) An OpenBCI Cyton Board (8 electrode channels) or an OpenBCI 16-channel R&D Kit (16 electrode channels)

image

####(1x) ~500mAh lithium ion rechargeable battery pack (Sparkfun or Adafruit) & (1x) A charger for your battery pack

image

SUGGESTED PRINT SETTINGS

If you're 3D-printing your Mark IV yourself, here are the print settings we recommend:

  • FRAME_FRONT & FRAME_BACK
    • Material: PLA
    • Supports: YES
    • Raft: hopefully NO (but if supports aren't sticking, try the raft)
    • Infill: 20%
    • Layer Height: 0.2mm
    • Number of Shells: 3
    • Speed while extruding: 50-70% (slow it down if possible; these parts are detailed)
  • MECH_PARTS (INSERT )
    • Material: PLA
    • Supports: NO
    • Raft: NO
    • Infill: 20%
    • Layer Height: 0.2mm
    • Number of Shells: 3
    • Speed while extruding: 50-70% (slow it down if possible; these parts are detailed)
  • BOARD_MOUNT, BOARD_COVER
    • Material: PLA
    • Supports: NO
    • Raft: NO
    • Infill: 20%
    • Layer Height: 0.2mm
    • Number of Shells: 3
    • Speed while extruding: 50-70% (slow it down if possible; these parts are detailed)
  1. Loctite Super Glue w/ Cyanoacrylate
  2. coarse flat & circular files (for removing support artifacts)
  3. medium sand paper
  4. exacto blade
  5. philips head screw driver
  6. wire cutters
  7. needle-nose pliers
  8. snippers

image

ASSEMBLY INSTRUCTIONS

Remove residual support material & print flaws

Use sand paper, a file, and snippers to clean your FRAME and other 3D-printed parts. The most important part of this process is that you thoroughly clean out the frames nodes where you will place your INSERTS pieces.

image image

Glue the FRAME together

Carefully, glue the FRONT FRAME and BACK FRAME together with your Cyanoacrylate super glue. The best way to do this is to place both halves of the frame on a level surface and carefully bring them together. Be sure to be precise; it's VERY difficult to pull the pieces apart once you've put them together.

image

Mount the OpenBCI BOARD_MOUNT

Use two screws to mount the Mark IV Board_Mount to the back (the more rounded half) of the Frame. Make sure that the orientation of the BOARD_HOLDER matches that of the pictures below:

image

Insert INSERTS pieces (x35) into frame

Before glueing the INSERTS into the frame, ensure that they fit properly into the frame without glue. INSERTs should be inserted from the inside of the frame outwards, so that they lie flush with the frame.

image

For each INSERT, line the inner rim of the frame with glue. Then insert the INSERT so that it is flush with the frame.

image

Screw 5 COMFORT UNITS into the frame as shown below. Your Ultracortex should now look like this:

image

Embed OpenBCI into the Ultracortex

Connect your ~500 mAh lithium ion rechargeable batter to the back of your Cyton OpenBCI Board:

image

Fold the battery and its wires neatly behind the board before inserting the board into the BOARD_MOUNT:

image

Snap your Cyton board onto the four pegs on the outside of the BOARD_MOUNT:

image

You can then secure the OpenBCI Board to the BOARD_MOUNT by clipping BOARD_COVER over top.

image

The pins and female headers should fit neatly into the BOARD_COVER holes as shown.

Only have a black battery pack with AA batteries? Tape the pack to your Mark IV in the location shown. As long as you use an 8-channel setup, the battery pack won't overlap with any of the electrode locations. Thread the battery pack wire through the hole in the BOARD_MOUNT as shown.

image

Connect your OpenBCI board to the battery pack, place it on the BOARD_MOUNT and snap the BOARD_COVER into place. Your Mark IV should look like the image below.

image

Electrode location overview

Before creating your electrode mounts, it's a good idea to think about where you may want to place the electrodes on the Ultracortex FRAME. The placement of the electrode may affect how long you make the wire between the electrode and where the OpenBCI is mounted, at the back of the frame.

The Ultracortex node locations are based on the 10-20 system, which is the internationally accepted standard for electrode placement in the context of EEG.

The images below indicates the default 10-20 electrode locations that the OpenBCI Graphical User Interface expects. This application is great for viewing/recording your EEG and can be found in our OpenBCI_Processing repo. The blue nodes indicate the 8 default 10-20 locations (channels 1-8) of the Cyton Board. The red nodes indicate the default 10-20 locations of channels 9-16 when using the OpenBCI 16-channel R&D Kit.

For the remainder of this tutorial, the blue nodes on the 10-20 system diagram (channels 1-8 of the OpenBCI default settings) will be used. The channel to 10-20 system correlations are as follows:

  • Channel 1(N1P) - Fp1
  • Channel 2(N2P) - Fp2
  • Channel 3(N3P) - C3
  • Channel 4(N4P) - C4
  • Channel 5(N5P) - P7
  • Channel 6(N6P) - P8
  • Channel 7(N7P) - O1
  • Channel 8(N8P) - O2

image

Placement of electrodes

First screw your two FLAT ELECTRODE UNITS in the front two nodes of the frame.

image

Then screw 6 SPIKEY ELECTRODE UNITS in the following nodes of the frame.

image

Connect wiring to OpenBCI

It's time to connect the electrodes to your OpenBCI Cyton board with jumper cables. You will find 3 ribbon cables in your kit as seen below.

image

Peel off the GRAY and PURPLE from the 12" set and the BLUE, GREEN, ORANGE and YELLOW from the 8" set and the RED and BROWN from the 4" Set as seen below.

image

Now connect the jumper cables to the electrodes. The map below assigns names to all the electrodes by their location. To follow along with this tutorial, attach each electrode to the wire color listed in the table below.

image

Electrode Wire Color
FP2 Grey
FP1 Purple
C4 Blue
C3 Green
P8 Yellow
P7 Orange
O2 Brown
O1 Red

To connect wires to electrodes, plug the male end in to the header cable on the NODES as shown:

image

After connecting wires to electrodes, run wires along the frame to the top of the OpenBCI BOARD_MOUNT. Use the included plastic clips to hold wires on the frame, like shown:

image image

Ear Clips

To use your Mark IV headset, you'll also need two ear clip electrodes. These ear clip electrodes serve as the reference and bias (ground with common-mode noise rejection) for your EEG system. You will connect them to your OpenBCI board along with the electrodes in the next step.

Cyton Board Setup

Connect an OpenBCI Cyton board to the Mark IV as shown below:

image

Here's a table with the correct electrode, wire color, and Cyton board pin pairings. By bottom pin, we mean the pin CLOSEST to the OpenBCI board:

Electrode Wire Color Cyton Board Pin
Ear Clip Black Bottom SRB pin (SRB2)
FP1 Purple Bottom N1P pin
FP2 Grey Bottom N2P pin
C3 Green Bottom N3P pin
C4 Blue Bottom N4P pin
P7 Orange Bottom N5P pin
P8 Yellow Bottom N6P pin
O1 Red Bottom N7P pin
O2 Brown Bottom N8P pin
Ear Clip Black Bottom BIAS pin

The electrodes (FP1 through O2) can be connected to any N1P, N2P, etc, channel. The ear clips must always be connected to the bottom SRB pin and either of the BIAS pins.

Replace the cover, as shown below:

image

Your Cyton board is ready to use with your Mark IV headset!

Ganglion Board Setup

Connect an OpenBCI Ganglion board to the Mark IV as shown below (with cover removed for clarity). We placed 8 electrodes on our Mark IV headset, but the Ganglion board can only take 4 inputs. So, we can only connect 4 of our electrodes to the board - we chose FP1, FP2,O1, and O2 for this tutorial.

image

Here's a table with the correct electrode, wire color, and Ganglion board pin pairings:

Electrode Wire Color Cyton Board Pin
FP2 Grey +4 (Top 4 pin)
FP1 Purple +3 (Top 3 pin)
O2 Brown +2 (Top 2 pin)
O1 Red +1 (Top 1 pin)
Ear Clip Black Top D_G pin
Ear Clip Black Top REF pin

Next, make sure the switches on your Ganglion board are in the bottom configuration as shown below:

Switches DOWN

Replace the cover, like below:

image

Your Ganglion board is now ready to use with your Mark IV headset!

Completed 8 Channel Mark IV Setup

Your 8 channel headset is finished! To expand to a 16 channel headset (for use with Cyton Daisy), see the next section.

image image image

16 Channel Add-Ons

If you have a Cyton Daisy extension, you can expand your headset from 8 electrodes to 16 electrodes. Each electrode provides one "channel" of data, so we call this a 16 channel setup.

Adding extra electrodes

From the front of the frame remove the following two Comfort Nodes.

image

Replace the old Comfort Nodes with spiky electrodes:

image

Add 6 more spiky electrode units in the locations shown below:.

image

Wiring Electrodes

Take out your rest of your Ribbon Cables and separate the GRAY and PURPLE 8" & 4" cables, the BLUE, GREEN, ORANGE and YELLOW 8", and the RED & BROWN 4" cables.

image

Plug the male end of the 8" GRAY and PURPLE to the 4" GRAY and PURPLE. These two extra-long wires will help you reach all the electrodes.

image

As shown below, connect one female end of the white Y-Splitter Cable to the bottom SRB pin of the Cyton. Connect the male end of the white Y-Splitter Cable to the female end of one black ear clip electrode. The other black ear clip electrode should be connected to the bottom BIAS pin of the Cyton. Bottom pins N1P through N8P of the Cyton should be connected to colored wires in the order shown below. Then, attach the Daisy extension to the outside of the Cyton board as shown below. You will plug all of the new electrodes into the Daisy board pins in the next step.

image

Connect each of the new electrodes to corresponding Daisy pins as shown below. You may have to use a different color scheme, depending on what wires you have available. Just make sure to keep track of which electrodes are connected to which Daisy pins.

image

The electrode and Daisy pin pairings are below:

Electrode Wire Color Daisy Pin
F7 Purple Bottom N1P pin
F8 Grey Bottom N2P pin
F3 White Bottom N3P pin
F4 Green Bottom N4P pin
T7 Brown Bottom N5P pin
T8 Red Bottom N6P pin
P3 Yellow Bottom N7P pin
P4 Orange Bottom N8P pin

The other female end of the white Y-Splitter Cable should be connected to the bottom SRB pin of the Daisy. The pins connected to the Daisy board should look like this:

image

Like for the first 8 electrodes, use the plastic clips to secure wires in place:

image

Your Mark IV is now ready to use 16 channels! It should look like the images below.

image image

Adjust the Ultracortex for your head

Put the Ultracortex Mark IV onto your head and gradually tighten the electrode units until the electrodes are snugly (but comfortably) against your scalp. Tighten the electrodes and comfort units by turning them clockwise, and loosen them by turning counterclockwise.

Be careful not to strain the electrode wires when twisting the electrode units, or you may separate the wire from the electrode itself.

Examine your brain waves!

Place your UltraCortex on your head so that the back center node is roughly the same distance above your inion (bump on the back of your skull) as the front center node is above the bridge of your nose. As you place the UltraCortex on your head, the springs should adjust to the shape and size of your head.

Now that you have your Ultracortex assembled and comfortably adjusted to your head size and shape, it's brain wave time!

image

Check out the Getting Started w/ OpenBCI tutorial to get up-and-running with the OpenBCI Processing GUI.

Below is a screenshot of what the GUI looks like when you've got your OpenBCI + Ultracortex (w/ 8 channels) hooked up! You can see a nice alpha (~11 hz) spike on the FFT Plot.

image

OpenBCI has a widget for visualizing signal strenth at each electrode. To view it, click on the "FFT Plot" drop-down menu in the top right corner. Select "Head Plot".

You should now see a visual map of signal strength in the top right corner.

image

Give Us Feedback!

If you have questions, comments, or suggestions regarding the printing and/or assembly of the Ultracortex, we'd love to hear from you. Please submit issues to this repository or email us at contact@openbci.com. Or take some pictures of yourself wearing the Ultracortex and Tweet at us (@OpenBCI & @Ultracortex)!